30 Keysight 16452A Liquid Test Fixture
Operation
Liquid Capacitance and Resistance (Cp, Rp) Measurement
3-
Liquid Capacitance and Resistance (C
p
, R
p
) Measurement
1. Pour the liquid under test into the 16452A until the liquid fills tube
Pour the liquid gently so that air does not get mixed in with the liquid.
Figure 3-9 Poring the Liquid under Test
Be careful when dealing with liquids (especially volatile and/or flammable
liquid such as oil or organic solvents) that can cause an accident.
Accidents such as an explosion, ignition, emission of poison gas, scalding
(by heat or chemicals), and so on, are possible.
To prevent any accidents, establish a strict process for the measurements.
(For example, using appropriate ventilation, a gas mask, and gloves.)
When you pour the liquid, be careful that the liquid does not leak from the
test fixture.
An 16452A assembled in the wrong manner (for example, lack of the
O-rings or the spacer) will cause liquid leakage. Assemble the 16452A
again if leakage is found.
If the problem cannot be corrected, please contact the nearest Keysight
Technologies sales office.
DO NOT enclose the liquid in the
16452A. Carburetion of the liquid
enclosed in the
16452A can cause an explosion of the 16452A.