42 Keysight Agile Signal Generator Service Guide
Troubleshooting
No Front Panel USB Activity
The front panel Type-A USB connectors support data transfers.
1. Connect a compatible USB memory stick to the signal generator. The
signal generator should detect the USB memory stick immediately.
2. If a compatible memory stick does not work, but the rear panel USB and
all other instrument functions operate correctly, replace the USB Board.
3. If replacing the USB Board does not correct the problem, replace the A7
CPU assembly.
Not all USB memory sticks are compatible with the Type-A USB connector on this signal
generator. Go to www.Keysight.com/find/x-series for a list of USB storage devices that work
with this signal generator.