80                                   Getting Started Guide
Test Platform Operating System
Updating the Keysight 5G NR Test Application
Scroll to find the 
Keysight E7515B application, 
and select 
Uninstall.
When prompted to completely remove the selected 
application and all of its features, select 
Yes.  
 
When the software uninstall is complete, select 
Finish.
The UXM 5G Control Panel should be closed at the beginning of 
the uninstallation. You must launch it manually at the end of this 
process.
3. Obtain the latest version of software 
installer.
Download it from the site: 
 
www.keysight.com/find/softwaremanager 
4. Locate the software installer program 
obtained in the step above.
5. Follow the on-screen prompts to 
continue the installation.
 
 
The installation process takes about 2 minutes. Do not turn the 
instrument power off or serious damage may occur. If any pop-up 
windows appear, click OK or Ignore to proceed. 
When the installation has finished, you must 
manually launch the UXM 5G Control Panel (from 
Desktop or Windows > Start).
6. Verify the UXM 5G is operational.  If the Control Panel shows a Faulty state, restart the UXM 5G or 
power cycle the UXM 5G as follows:  
a. Power off the UXM 5G and disconnect the power cord for 15 
seconds.  
b. Power on the UXM 5G.
Step Notes