52 M8100 Series Arbitrary Waveform Generators Getting Started Guide
2 Basic Setup for M8100A
 
Case 3 Connecting M8008A clock module to four M8199B AWG modules
In this case, a M8008A clock module, four M8199B AWG modules, two 
M9505A AXIe chassis and the following cables are required:
• 3.5 mm, 50 cm M8199A-811 customer orderable sync cable 
(M8199-61620)
• 1.85 mm, 45 cm M8199A-810 replacement channel clock cable 
(M8199-61624)
Make the connections as described below:
1 Connect the M8008A Clock Generator Sync Out port with all the four 
M8199B AWG Sync In ports using sync cable (blue cable).
2 Connect the M8008A Clock Generator Sample Clock Out port with all 
the four M8199B AWG Clock In ports using clock cable (red cable).
Figure 22 shows how to connect M8008A clock generator module to the 
four M8199B AWG modules: