230 Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
13 Acquisition Control
If the signal you are measuring has more than about 1 s of inactivity, consider
selecting the Normal trigger mode to prevent AutoTriggering. See “To select the
Auto or Normal trigger mode" on page 208.
See Also “Navigating Segments" on page 230
“Measurements, Statistics, and Infinite Persistence with Segmented
Memory" on page 231
“Segmented Memory Re-Arm Time" on page 231
“Saving Data from Segmented Memory" on page 231
Navigating Segments
1 Press the Current Seg softkey and turn the Entry knob to display the desired
segment along with a time tag indicating the time from the first trigger event.
You can also navigate segments using the [Navigate] key and controls. See “To
navigate segments" on page 76.