295
Keysight InfiniiVision 3000T X-Series Oscilloscopes 
User's Guide
 
18 Frequency Response 
Analysis
To make connections / 295
To set up and run the analysis / 296
To view and save the analysis results / 298
Frequency Response Analysis (FRA) is a licensed feature you can get by 
purchasing the DSOXT3FRA upgrade.
The Frequency Response Analysis (FRA) feature controls the built-in waveform 
generator to sweep a sine wave across a range of frequencies while measuring the 
input to and output from a device under test (DUT). At each frequency, gain (A) 
and phase are measured and plotted on a frequency response Bode chart.
When the frequency response analysis completes, you can move a marker across 
the chart to see the measured gain and phase values at each frequency point. You 
can also adjust the chart's scale and offset settings for the gain and phase plots.
To make connections
The waveform generator output is connected to a device under test (DUT). The 
input to the device and the output from the device are probed by the 
oscilloscope's input channels.