Pulse
Back to ARB Commands
[SOURce:]ARB:CURRent:PULSe:END:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:PULSe:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:PULSe:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:PULSe:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:PULSe:END:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time in seconds, after the pulse completes, that the starting current, voltage, power, or
resistance level persists.
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:PULSe:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:END:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:PULSe:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:END:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:PULSe:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:END:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:PULSe:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time in seconds, after the pulse completes, that the starting current, voltage, power, or
resistance level persists for Arb sequence steps.
Parameter Typical Return
0 – 262.144 | MIN | MAX, *RST 0 <end time>
Step 0 - 99, *RST 0 (none)
Programs an end time for the current pulse: ARB:CURR:PULS:END:TIM 2,(@1)
Programs an end time for the voltage pulse:ARB:VOLT:PULS:END:TIM 2,(@1)
Programs an end time for the power pulse:ARB:POW:PULS:END:TIM 2,(@1)
Programs an end time for the resistance pulse:ARB:RESISTANCE:PULS:END:TIM 2,(@1)
Programs an end time for step 0: ARB:SEQ:STEP:CURR:PULS:END:TIM 2, 0,(@1)
5 SCPI Programming Reference
194 Keysight N6700C Operating and Service Guide