[SOURce:]ARB:CURRent:RAMP:END:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:RAMP:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:RAMP:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:RAMP:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:RAMP:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:RAMP:END:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time in seconds, after the ramp completes, that the ending current, voltage, power, or
resistance level persists.
[SOURce:]ARB:SEQuence:STEP:CURRent:RAMP:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:RAMP:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:RAMP:END:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:RAMP:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:RAMP:END:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:RAMP:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:RAMP:END:TIMe<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:RAMP:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time in seconds, after the ramp completes, that the ending current, voltage, power, or
resistance level persists for Arb sequence steps.
Parameter Typical Return
0 262.144 | MIN | MAX, *RST 0 <end time>
Step 0 - 99, *RST 0 (none)
Programs an end time for the current ramp: ARB:CURR:RAMP:END:TIM 2,(@1)
Programs an end time for the voltage ramp:ARB:VOLT:RAMP:END:TIM 2,(@1)
Programs an end time for the power ramp:ARB:POW:RAMP:END:TIM 2,(@1)
Programs an end time for the resistance ramp: ARB:RESISTANCE:RAMP:END:TIM 2,(@1)
Programs an end time for step 0: ARB:SEQ:STEP:CURR:RAMP:END:TIM 2, 0,(@1)
[SOURce:]ARB:CURRent:RAMP:RTIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:RAMP:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:RTIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:RAMP:RTIMe<value>, (@<chanlist>)
[SOURce:]ARB:POWer:RAMP:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:RAMP:RTIMe<value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:RAMP:RTIMe? [MIN|MAX,] (@<chanlist>)
Specifies the rise time of the ramp in seconds.