Sinusoid
Back to ARB Commands
[SOURce:]ARB:CURRent:SINusoid:AMPLitude<value>, (@<chanlist>)
[SOURce:]ARB:CURRent:SINusoid:AMPLitude? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:AMPLitude <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:AMPLitude? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:POWer:SINusoid:AMPLitude <value>, (@<chanlist>)
[SOURce:]ARB:POWer:SINusoid:AMPLitude? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:RESISTANCE:SINusoid:AMPLitude <value>, (@<chanlist>)
[SOURce:]ARB:RESISTANCE:SINusoid:AMPLitude? [MIN|MAX,] (@<chanlist>)
Specifies the amplitude or peak value of the current, voltage, power, or resistance sine wave.
[SOURce:]ARB:SEQuence:STEP:CURRent:SINusoid:AMPLitude<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:SINusoid:AMPLitude? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:SINusoid:AMPLitude <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:SINusoid:AMPLitude? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:SINusoid:AMPLitude <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:SINusoid:AMPLitude? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:SINusoid:AMPLitude <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:SINusoid:AMPLitude? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the amplitude or peak value of the current, voltage, power, or resistance sine wave for Arb
sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
Power: 0 - maximum|MIN|MAX, *RST MIN
Resistance: 0 - maximum|MIN|MAX, *RST MIN
Values are range and module-rating dependent.
<amplitude>
Step 0 - 99, *RST 0 (none)
Programs a current amplitude: ARB:CURR:SIN:AMPL 5,(@1)
Programs a voltage amplitude: ARB:VOLT:SIN:AMPL 10,(@1)
Programs a power amplitude: ARB:POW:SIN:AMPL 30,(@1)
Programs a resistance amplitude: ARB:RESISTANCE:SIN:AMPL 50,(@1)
Programs a current amplitude for step 0: ARB:SEQ:STEP:CURR:SIN:AMPL 5, 0,(@1)
Keysight N6700C Operating and Service Guide 203
5 SCPI Programming Reference