Parameter Typical Return
Current: 0 - maximum
Voltage: 0 - maximum
Power: 0 - maximum
Resistance: 0 - maximum
Values are range and module-rating dependent. *RST 1 point setto
the minimum value.
<value> [,<value>]
Step 0 - 99, *RST 0 (none)
Program the current points: ARB:CURR:UDEF5,4,3,2,1,(@1)
Program the voltage points: ARB:VOLT:UDEF 20,21,22,23,24,(@1)
Program the power points: ARB:POW:UDEF 20,30,40,50,60,(@1)
Program the resistance points: ARB:RESISTANCE:UDEF 60,70,80,90,100,(@1)
Program the points for step 0: ARB:SEQ:STEP:CURR:UDEF 5,4,3,2,1,0,(@1)
l At the end of the user-defined Arb, the output state of the unit depends upon the ARB:TERMin-
ate:LAST program settings.
l The order in which the values are entered determines the sequence of execution.
l To create a valid user-defined Arb, the current level, voltage level, power level, resistance level,
BOST, and dwell lists must either all be the same length, or have a length of 1, which is inter-
polated as having the same length as the maximum length list.
[SOURce:]ARB:CURRent:UDEFined:LEVel:POINts? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:LEVel:POINts? (@<chanlist>)
[SOURce:]ARB:POWer:UDEFined:LEVel:POINts? (@<chanlist>)
[SOURce:]ARB:RESISTANCE:UDEFined:LEVel:POINts? (@<chanlist>)
Returns the number of current, voltage, or resistance points in the user-defined Arb, not the point
values.
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:LEVel:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:LEVel:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:UDEFined:LEVel:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:UDEFined:LEVel:POINts? <step#>, (@<chanlist>)
Returns the number of current, voltage, power, or resistance points in the user-defined Arb for Arb
sequence steps.
Parameter Typical Return
(none) <number of points>
Step 0 - 99, *RST 0 (none)
Returns the number of current points: ARB:CURR:UDEF:LEV:POIN? (@1)
Returns the number of voltage points: ARB:VOLT:UDEF:LEV:POIN? (@1)
Returns the number of power points: ARB:POW:UDEF:LEV:POIN? (@1)
Returns the number of resistance points: ARB:RESISTANCE:UDEF:LEV:POIN? (@1)
Returns the number of points for step 0: ARB:SEQ:STEP:CURR:UDEF:LEV:POIN? 0,(@1)
5 SCPI Programming Reference
224 Keysight N6700C Operating and Service Guide