Parameter Typical Return
Programs a current list of 3 steps. LIST:CURR 3,2,1, (@1)
Programs a voltage list of 3 steps. LIST:VOLT 20,10,5, (@1)
Programs a power list of 3 steps. LIST:POW30,40,50, (@1)
Programs a resistance list of 3 steps. LIST:RES 20,50,100, (@1)
l The order in which the list values are entered determines the sequence when the list executes.
l To create a valid list, the current, voltage, power, resistance, BOST, EOST, and dwell lists must all
be the same length or have a length of 1, which is interpreted as having the same length as the
longest list.
l This command overwrites any present list in active memory; it does not append to the present list.
[SOURce:]LIST:DWELl <value>{,<value>}, (@<chanlist>)
[SOURce:]LIST:DWELl? (@<chanlist>)
Specifies the dwell time for each list step. Dwell time is the time that the output will remain at a
specific step. Dwell times can be programmed from 0 through 262.144 or 0 through 268.435 seconds
with the following resolution:
All models except N678xA SMU and N679xA
Models N678xA SMU and N679xA only
Range in seconds Resolution
Range in seconds
Resolution
0 - 0.262144 1 microsecond
0 - 0.268435
1.024 microsecond
0.262144 - 2.62144 10 microseconds
0.268435 - 2.68435
10.24 microseconds
2.62144 - 26.2144 100 microseconds
2.68435 - 26.8435
102.4 microseconds
26.2144 - 262.144 1 millisecond
26.8435 - 268.435
1.024 millisecond
A comma-delimited list of up to 512 steps may be programmed. At the end of the dwell time, the
output state of the unit depends upon the [SOURce:]LIST:STEP program settings.
Parameter Typical Return
0 262.144 (all except N678xA and N679xA)
0 - 268.435 (N678xA SMU and N679xA only)
*RST 1 step with a value of 0.001
<value 1>, <value 2>, <value 3>...
Programs a dwell list. The list contains 3 steps: LIST:DWEL 0.2,0.8,1.6, (@1)
l The order in which the list values are entered determines the sequence when the list executes.
l To create a valid list, the current, voltage, power, resistance, BOST, EOST, and dwell lists must all
be the same length or have a length of 1, which is interpreted as having the same length as the
longest list.
l This command overwrites any present list in active memory; it does not append to the present list.
Keysight N6700C Operating and Service Guide 257
5 SCPI Programming Reference