[SOURce:]LIST:CURRent:POINts? (@<chanlist>)
[SOURce:]LIST:DWELl:POINts? (@<chanlist>)
[SOURce:]LIST:VOLTage:POINts? (@<chanlist>)
[SOURce:]LIST:POWer:POINts? (@<chanlist>)
[SOURce:]LIST:RESistance:POINts? (@<chanlist>)
[SOURce:]LIST:TOUTput:BOSTep:POINts? (@<chanlist>)
[SOURce:]LIST:TOUTput:EOSTep:POINts? (@<chanlist>)
Returns the number of list points (same as steps). The queries do not return the point values.
Parameter Typical Return
(none) <points>
Returns the number of points in the current list: LIST:CURR:POIN? (@1)
Returns the number of points in the dwell list: LIST:DWEL:POIN? (@1)
Returns the number of points in the voltage list: LIST:VOLT:POIN? (@1)
Returns the number of points in the power list: LIST:POW:POIN? (@1)
Returns the number of points in the resistancelist: LIST:RES:POIN? (@1)
Returns the number of points in the BOST list: LIST:TOUT:BOST:POIN? (@1)
Returns the number of points in the EOSTlist: LIST:TOUT:EOST:POIN? (@1)
[SOURce:]LIST:STEP ONCE|AUTO, (@<chanlist>)
[SOURce:]LIST:STEP? (@<chanlist>)
Specifies how the list responds to triggers.
ONCE - the output remains at the present step until a trigger advances it to the next step. Triggers
that arrive during the dwell time are ignored.
AUTO - the output automatically advances to each step, after the receipt of an initial starting trigger.
Steps are paced by the dwell list. As each dwell time elapses, the next step is immediately output.
Parameter Typical Return
ONCE|AUTO, *RST AUTO ONCE or AUTO
Specifies the list steps to be paced by trigger signals: LIST:STEP ONCE, (@1)
[SOURce:]LIST:TERMinate:LAST 0|OFF|1|ON, (@<chanlist>)
[SOURce:]LIST:TERMinate:LAST? (@<chanlist>)
Determines the output value when the list terminates. When ON (1), the output remains at the last list
step. The value of the last list step becomes the IMMediate value. When OFF (0), and also when the
list is aborted, the output returns to the settings that were in effect before the list started.
Parameter Typical Return
0|OFF|1|ON , *RST OFF 0or1
Terminate with the output at the last step value: LIST:TERM:LASTON, (@1)
5 SCPI Programming Reference
258 Keysight N6700C Operating and Service Guide