Downprogrammer Calibration
Step 1. Remove all loads from the output. This procedure is automatic and takes a few seconds.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Misc\Dprog CAL:DPRog (@1)
Enter a Calibration Date, Save, and Log Out
Front Panel Menu Reference SCPI Command
Select System\Admin\CAL\DATE.
Enter the calibration date in the Datefield.
Select System\Admin\CAL\Save.
Select Save to save all calibration data.
Select System\Admin\Logout
to exitcalibration mode.
To enter a calibration date:
CAL:DATE "<date>"
To save the calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
Keysight N6700C Operating and Service Guide 359
6 Verification and Calibration