Front Panel Menu Reference SCPI Command
Display shows: “Enter P3 measured data”
Enter the data from the external DMM. Press Select when done. Press Next to
finish calibration.
CAL:LEV P3
*OPC?
CAL:DATA <data>
Step 6. For Model N6781A repeat steps 1 through 5 for the 6 V resistance range. This time, select the
Low range.
Enter a Calibration Date, Save, and Log Out
Front Panel Menu Reference SCPI Command
Select System\Admin\CAL\DATE.
Enter the calibration date in the Datefield.
Select System\Admin\CAL\Save.
Select Save to save all calibration data.
Select System\Admin\Logout
to exitcalibration mode.
To enter a calibration date:
CAL:DATE "<date>"
To save the calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
Keysight N6700C Operating and Service Guide 369
6 Verification and Calibration