Test Record Form - Keysight N6751A and N6752A
N6751A and N6752A Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
Constant Voltage Tests
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
− 1 mV
Vout − 20 mV
49.951 V
Vout − 45 mV
_______
_______
_______
_______
+ 39 mV
Vout + 20 mV
50.049 V
Vout + 45 mV
CV Load Effect: Both – 2 mV _______ + 2 mV
CV Source Effect: Both – 1 mV _______ + 1 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 4.5 mV
+ 0.35 mV
Transient response
Voltage @ 100 μs:
Voltage @ 100 μs with Option 761:
Both
N6752A
– 75 mV
– 125 mV
_______
_______
+ 75 mV
+ 125 mV
Constant Current Tests
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
N6751A
N6752A
N6751A
N6752A
− 10 mA
Iout − 4 mA
4.975 A
9.970 A
Iout − 9 mA
Iout − 14 mA
_______
_______
_______
_______
_______
_______
+ 30 mA
Iout + 4 mA
5.025 A
10.030 A
Iout + 9 mA
Iout + 14 mA
CC Load Effect: Both – 2 mA _______ + 2 mA
CC Source Effect: Both – 1 mA _______ + 1 mA
Test Description N6751A Setting N6752A Setting
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
CV Load Effect, Source Effect, Ripple and Noise:
Transient Response:
Current Programming & Readback Min Current:
Current Programming & Readback High Current:
CC Load Effect, Source Effect:
20 mV, 1 A
50 V, 1 A
50 V, 1 A
10 V, from 3 A to 5 A
10 mA, 10 V
5 A, 10 V
5 A, 10 V
20 mV, 2 A
50 V, 2 A
50 V, 2 A
8.5 V, from 5 A to 10 A
10 mA, 8.5 V
10 A, 8.5 V
10 A, 8.5 V
6 Verification and Calibration
384 Keysight N6700C Operating and Service Guide