Feature
(l = available)
Source/Measure Units (SMU) Application-Specific
N6781A N6782A N6784A N6785A N6786A N6783A
-BAT
N6783A
-MFG
Output rating 20 W 20 W 20 W 80 W 80 W 24 W 18 W
2-quadrant operation
l l l l l l
4-quadrant operation
l
Auxiliary voltage input
l l
Output disconnect relays
l l l l l
Opt. 761 Opt. 761
Arbitrary waveform generation
1
l l l l l l l
Negative voltage protection
l l l l l l l
Voltage or current priority
l l l l l
Programmable resistance
l l
Number of voltage output ranges 3 3 3 4 4 1 1
Number of current output ranges 3 3 4 4 4 1 1
Number of voltage measurement ranges 3 3 3 1 1 1 1
Number of current measurement ranges 4 4 4 3 3 2 2
Simultaneous voltage and current meas-
urements
l l l l l
Seamless measurement autoranging
l l
Opt. SMR
l l
SCPI output list capability
1, 2
l l l l l l l
SCPI array readback
2
l l l l l l l
SCPI programmable sample rate
2
l l l l l l l
SCPI external data logging
2
l l l l l l l
Double-wide (occupies two channel locations)
l l
Note 1 Arbitrary waveforms and List capability is not available on the negative current output on Model N6783A
Note 2 Only available when using the remote interfaces; not from the front panel.
1 Quick Reference
42 Keysight N6700C Operating and Service Guide