3 Spectrum Analyzer Mode
3.8 Spurious Emissions Measurement
n Return Value
1 or omitted Returns a variable-length (1 + 6 * Spurs up to 1201 entries) comma-separated list containing detailed
information in the following format:
Number of spurs in following list (Integer)
[Repeat the following for each spur]
Spur #
Range # Spur was located (Integer)
Frequency of Spur (Hz, Float64)
Amplitude of Spur (dBm, Float32)
Absolute Limit (dBm, Float32)
Pass or Fail (1|0, Boolean)
2 21 (Average
Trace)
Regardless of the Trace selection, returns a comma-separated list of the average trace data for the
selected range (where range number = n 1) using Detector 1. If selected range is not active, SCPI_
NAN is returned for each trace data element
22 Returns the number of spurs found for the selected Measured Trace
23 42 (Average
Trace)
Regardless of the Trace selection, returns a comma-separated list of the average trace data for the
selected range (where range number = n 22) using Detector 2. If selected range is not active or
Detector 2 selection is off, SCPI_NAN is returned for each trace data element
43 62
(Maximum Hold
Trace)
Regardless of the Trace selection, returns a comma-separated list of the maximum hold trace data for
the selected range (where range number = n 42) using Detector 1. If selected range is not active,
SCPI_NAN is returned for each trace data element
63 82
(Minimum Hold
Trace)
Regardless of the Trace selection, returns a comma-separated list of the minimum hold trace data for
the selected range (where range number = n 62) using Detector 1. If selected range is not active,
SCPI_NAN is returned for each trace data element
83-102 Reserved
103-122 Reserved
123-142
(Clear/Write
Trace)
Regardless of the Trace selection, returns a comma-separated list of the clear/write trace data for the
selected range (where range number = n 122) using Detector 1. If selected range is not active,
SCPI_NAN is returned for each trace data element
143-162
(Clear/Write
Trace)
Regardless of the Trace selection, returns a comma-separated list of the clear/write trace data for the
selected range (where range number = n 142) using Detector 2. If selected range is not active or
Detector 2 selection is off, SCPI_NAN is returned for each trace data element
163-182 (Trace
1)
Returns a comma-separated list of the trace data of Trace 1 for the selected range (where range
number = n 162). If selected range is not active, SCPI_NAN is returned for each trace data element
183-202 (Trace
2)
Returns a comma-separated list of the trace data of Trace 2 for the selected range (where range
number = n 182). If selected range is not active, SCPI_NAN is returned for each trace data element
203-222 (Trace
3)
Returns a comma-separated list of the trace data of Trace 3 for the selected range (where range
number = n 202). If selected range is not active, SCPI_NAN is returned for each trace data element
223 Returns "Marker Table" on page 1358 data as a series of comma separated values in the following form: