3 Spectrum Analyzer Mode
3.9 SEM Measurement
Sweep Type
Sets the Sweep Type used to calculate the power in the reference channel. Sweep
Type can be set manually or put into Auto mode.
How to define Channel Sweep Time and Channel Sweep Type:
Channel Sweep Type
Mode
Behavior
Auto Channel Sweep Type is automatically selected depending on Sweep Type
Rules
Channel Sweep Time is automatically calculated depending on the
selected sweep type
Man Channel Sweep Type is user-selected
Remote Command
[:SENSe]:SEMask:SWEep[1]|2:TYPE SWEep | FFT
[:SENSe]:SEMask:SWEep[1]|2:TYPE?
[:SENSe]:SEMask:SWEep[1]|2:TYPE:AUTO OFF | 0 | ON | 1
[:SENSe]:SEMask:SWEep[1]|2:TYPE:AUTO?
Example
:SEM:SWE:TYPE FFT
:SEM:SWE:TYPE?
:SEM:SWE:TYPE:AUTO OFF
:SEM:SWE:TYPE:AUTO?
Notes Sub op code, 1 is for BTS, 2 for MS. Default is BTS
Note that Sweep sub op code 2 is supported only in non-SA Modes. In SA Mode, Sweep sub op code 1
is used for both BTS and MS
Dependencies Grayed-out in VXT models M9410A/11A
Couplings If Sweep Type is set manually, Sweep Type mode is set to MANual
When Channel Sweep Type mode is Auto, Sweep Type is automatically selected according to Sweep
Type Rules
Preset Automatically calculated
ON
State Saved Saved in instrument state
Yes
Range
OFF|ON
1694 Spectrum Analyzer Mode User's &Programmer's Reference