Loading...
9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Operation Instrument Negative Transition
Determines which bits in the "Operation Condition Query" on page 2788 Register
will set the corresponding bit in the "Operation Event Query" on page 2789 register
when the condition register bit has a negative transition (1 to 0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:OPERation:INSTrument:NTRansition <integer>
:STATus:OPERation:INSTrument:NTRansition?
Example
:STAT:OPER:INST:NTR 1
Instrument Locked being cleared will be reported to the Instrument Summary of the Status Operation
register
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Operation Instrument Positive Transition
Determines which bits in the "Operation Condition Query" on page 2788 Register
will set the corresponding bit in the "Operation Event Query" on page 2789 register
when the condition register bit has a positive transition (0 to 1).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:OPERation:INSTrument:PTRansition <integer>
:STATus:OPERation:INSTrument:PTRansition?
Example
:STAT:OPER:INST:PTR 1
Instrument Locked being set will be reported to the Instrument Summary of the Status Operation
register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
2793 Spectrum Analyzer Mode User's &Programmer's Reference