3 Spectrum Analyzer Mode
3.11 Harmonics Measurement
REPeat After reaching the "Avg/Hold Num" on page 1919, all previous result
data is cleared, and the average count set back to 1
Remote
Command
[:SENSe]:HARMonics:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:HARMonics:AVERage:TCONtrol?
Example
:HARM:AVER:TCON EXP
Preset
REPeat
Range
EXPonential|REPeat
Meas Setup Summary Table
Lets you view and access many of the parameters in the Meas Setup menus on one
screen.
Harmonics
Sets the number of harmonics that are measured, including the fundamental, thus,
setting this to 3 measures up through the third harmonic. All measured harmonics
are included in the calculation of Total Harmonic Distortion (THD).
Harmonics outside of the frequency range of the instrument are not measured.
Harmonics that require a resolution bandwidth greater than the instrument’s range
are measured using the widest available resolution bandwidth.
Remote Command
[:SENSe]:HARMonics:NUMBer <integer>
[:SENSe]:HARMonics:NUMBer?
Example
:HARM:NUMB 5
Preset 10
Min 2
Max 10
Range Table (On/Off)
Enables override of the default harmonic measurement parameters. Allows you to
set frequency, bandwidth, sweep time, etc. for each harmonic independently.
When Range Table is OFF, the data in the range table is cleared.
Remote
Command
[:SENSe]:HARMonics:RTABle:STATe OFF | ON | 0 | 1
[:SENSe]:HARMonics:RTABle:STATe?
Example
:HARM:RTAB:STAT 1
Spectrum Analyzer Mode User's &Programmer's Reference 1911