EasyManua.ls Logo

Keysight N9000B - LNA

Keysight N9000B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Spectrum Analyzer Mode
3.13 Phase and Amplitude vs Time Measurement
Example
:POW:GAIN:BAND LOW
:POW:GAIN:BAND?
Dependencies Not available on all hardware platforms. If the preamp is not present or is unlicensed, this control is
not shown
Does not appear in VXT Models M9410A/11A/15A/16A nor in M9410E/11E/15E/16E
If :POW:GAIN:BAND FULL is sent when a low band preamp is available, the preamp band
parameter is set to LOW instead of FULL, and an "Option not installed" message is generated
Not available when the electronic/soft attenuator is enabled
Preset
LOW
State Saved Saved in instrument state
Annotation When the USB Preamp is not connected to USB, the Preamp annotation in the Meas Bar says “Off” if
the preamp is off and displays the frequency range of the low band or full range preamp depending on
the setting. For example, if the Preamp is set to Low Band and the Low Band is 3.6GHz the annotation
says “3.6GHz”. If it is a 13.6GHz preamp and it is set to Full Range the annotation says “13.6GHz”
When the USB Preamp is connected to USB, the Preamp annotation says “Preamp: USB” if the
internal preamp is off or “Preamp: USB, Int” if the internal preamp is on (only for measurements that
support the USB preamp)
Auto Function
Remote Command
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF | ON | 0 | 1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Example
:POW:GAIN OFF
:POW:GAIN?
Preset
OFF
LNA
Lets you turn the Low Noise Amplifier (LNA) on or off.
LNA is an additional preamplifier that provides superior DANL and frequency range
compared to "Internal Preamp" on page 1981. LNA provides lower system noise
figure, especially at frequencies above 100MHz, and can be operated up to the full
range of 50GHz instruments.
For best possible sensitivity, LNA can be turned on together with "Internal Preamp"
on page 1981, although if you operate both preamps together, note that the TOI
(distortion) specifications are impacted. The sensitivity improvement of this
combination is substantial when operating in high band (frequencies above
3.6GHz).
For more details about annotation, see "More Information" on page 1983
Remote Command
[:SENSe]:POWer[:RF]:GAIN:LNA[:STATe] OFF | ON | 0 | 1