EasyManua.ls Logo

Keysight N9000B - Ref Position

Keysight N9000B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Spectrum Analyzer Mode
3.9 SEM Measurement
The on/off switch turns Ref Level Offset on or off. Setting a value for Ref Level Offset
turns it ON.
For more on using offsets, see "Reference Level Offset" on page 210.
Remote Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet?
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet:STATe OFF | ON | 0 | 1
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet:STATe?
Example Set the Ref Level Offset to 12.7dB. The only valid suffix is dB. If no suffix is sent, dB will be assumed:
:DISP:WIND:TRAC:Y:RLEV:OFFS 12.7
:DISP:WIND:TRAC:Y:RLEV:OFFS?
:DISP:WIND:TRAC:Y:RLEV:OFFS:STAT ON
Turns the Ref Level Offset On
Dependencies This control appears only in Spectrum Analyzer Mode
Preset 0 dBm
OFF
State Saved Saved in instrument state
Min The range for Ref Lvl Offset is variable. It is limited to values that keep the reference level within the
range of -327.6 dB to 327.6 dB
Max 327.6 dB
Annotation The offset is displayed as “Ref Offset <value>” to the right of the reference level annotation if nonzero.
When the offset is zero, no annotation is shown
Ref Position
Positions the reference level at the top, center, or bottom of the Y Scale display.
Changing the reference position does not change "Ref Value" on page 1555.
Remote
Command
:DISPlay:SEMask:WINDow[1]:TRACe:Y[:SCALe]:RPOSition TOP | CENTer | BOTTom
:DISPlay:SEMask:WINDow[1]:TRACe:Y[:SCALe]:RPOSition?
Example
:DISP:SEM:WIND:TRAC:Y:RPOS CENT
:DISP:SEM:WIND:TRAC:Y:RPOS?
Preset
TOP
State Saved Saved in instrument state
Range
TOP|CENTer|BOTTom
Annotation The greater than (>) and less than (<) symbols are displayed on both sides of the graph to indicate the
Reference Position
Backwards
Compatibility
SCPI
:DISPlay:SEMask:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RPOSition