3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
Remote
Command
:CALCulate:LLINe[1]|2|…|6:OFFSet:UPDate
Example Set updates the limit table to reflect the X and Y offsets, then reset the offsets to zero:
:CALC:LLIN:OFFS:UPD
State Saved No state
Delete Row
This is an immediate action key. It immediately deletes the currently-selected row,
whether or not that row is being edited. The row following the currently-selected
row (or the row preceding if there is none) is selected.
Delete Limit
Deletes the currently selected limit line. Pressing Delete Limit purges the data from
the limit line tables.
Limit data – including secondary parameters such as description, margin value, etc.
– is cleared and returned to factory preset settings.
When this control is pressed, a prompt is placed on the screen that says “Please
press Enter or OK to delete limit. Press ESC or Cancel to close this dialog.” The
deletion is only performed if you press OK or Enter. Following deletion, the
informational message Limit deleted appears in the MSG line.
Remote Command
:CALCulate:LLINe[1]|2|…|6:DELete
Example Delete all data for Limit Line 2:
:CALC:LLIN2:DEL
Limit Table
Lets you configure the properties of the current row of the Limit Table.
Frequency
Lets you edit the frequency of the current row.
Min 0
Max 1 THz
Amplitude
Lets you edit the Amplitude of the current row.
394 Spectrum Analyzer Mode User's &Programmer's Reference