6 Input/Output
6.10 Calibrator Control
Preset Unaffected by Mode Preset. Set to OFF by Restore Input/Output Defaults or Restore System Defaults-
>All
6.10.3 Cal Frequency
Lets you set the selected calibrator’s frequency.
Remote
Command
:SYSTem:CALibration:TUNE:FREQuency <freq>
:SYSTem:CALibration:TUNE:FREQuency?
Example Set source frequency to 150 MHz:
:SYST:CAL:TUNE:FREQ 150000000
Preset Unaffected by Mode Preset. Set to 1GHz by Restore Input/Output Defaults or Restore System
Defaults->All
Min/Max Depend on the selected calibrator
6.10.4 Cal Signal Type
Lets you set the selected calibrator’s signal type.
Remote Command
:SYSTem:CALibration:TUNE:TYPE CW | COMB
:SYSTem:CALibration:TUNE:TYPE?
Example
:SYST:CAL:TUNE:TYPE CW
:SYST:CAL:TUNE:TYPE?
Dependencies If the selected calibrator does not support a signal type, then that type is disabled in the dropdown
Changing the signal type to a disabled option generates an error
Preset Unaffected by Mode Preset. Set to CW by Restore Input/Output Defaults or Restore System Defaults-
>All
6.10.5 Cal Comb Spacing
Lets you set the calibrator’s comb spacing, when the signal type is COMB.
Remote Command
:SYSTem:CALibration:TUNE:SPACing <freq>
:SYSTem:CALibration:TUNE:SPACing?
Example Set comb spacing to 1 MHz:
:SYST:CAL:TUNE:SPAC 1000000
Dependencies Only appears when COMB is selected as "Cal Signal Type" on page 2521
If the selected calibrator does not support the Comb signal, attempting to set the spacing generates
an error
Spectrum Analyzer Mode User's &Programmer's Reference 2521