3 Spectrum Analyzer Mode
3.10 TOI Measurement
knob, and it behaves as it normally would in manual mode
The Auto/Man state of (Mech) Atten is saved
The Auto/Man toggle on the (Mech) Atten control disappears, and the auto rules
are disabled
The Electronic Attenuator is set to 10dB less than the previous value of the
Mechanical Attenuator, within the limitation that it must stay within the range of
0 to 24dB of attenuation
Examples in the Dual-Attenuator configuration:
Mech Atten at 20dB. Elec Atten enabled, Mech Atten set to 10dB, and Elec
Atten set to 10dB. New total attenuation equals the value before Elec Atten
enabled
Mech Atten at 0dB. Elec Atten enabled, Mech Atten set to 10dB, and Elec Atten
set to 0dB. New total attenuation does not equal the value before Elec Atten
enabled
Mech Atten at 40dB. Elec Atten enabled, Mech Atten set to 10dB, and Elec
Atten set to 24dB. New total attenuation does not equal the value before Elec
Atten enabled
When the Electronic Attenuation is disabled from an enabled state:
The Elec Atten control is grayed out
The Auto/Man state of (Mech) Atten is restored
If now in Auto, (Mech) Atten recouples
If now in Man, (Mech) Atten is set to the value of total attenuation that existed
before the Elec Atten was disabled. The resulting value is rounded up to the
smallest value possible given the (Mech) Atten Step setting - (That is, 57dB
changes to 58dB when (Mech) Atten Step is 2dB)
Using the Electronic Attenuator: Pros and Cons
The electronic attenuator offers finer steps than the mechanical attenuator, has no
acoustical noise, is faster, and is less subject to wear.
The “finer steps” advantage of the electronic attenuator is beneficial in optimizing
the alignment of the instrument dynamic range to the signal power in the front panel
as well as remote use. Thus, you can achieve improved relative signal measurement
accuracy. Compared to a mechanical attenuator with 2dB steps, the 1dB resolution
of the electronic attenuator only gives better resolution when the odd-decibel steps
are used. Those odd-decibel steps are less accurately calibrated than the even-
1776 Spectrum Analyzer Mode User's &Programmer's Reference