7 Save/Recall/Print
7.3 Save
Similarly, the Bias rows can only be used in Correction register 1, because there can
only be one setting for Bias at any given time. If a Correction file with a Bias or Bias
State row is loaded into any Correction register but 1, an error is generated: Mass
storage error; Can only load Bias Settings into Correction 1
The data follows the DATA row, as comma separated X, Y pairs; one pair per line.
For example, suppose you have an Antenna to correct for on an N9020A version
A.02.06 and the correction data is:
0dB at 200MHz
17dB at 210MHz
14.8dB at 225MHz
Then the file will look like:
Amplitude Correction
"Correction Factors for 11966E"
"Class B Radiated"
A.02.06,N9020A
P13 EA3 UK6,01
Frequency Unit, MHz
Antenna Unit, dBuV/m
Frequency Interpolation, Linear
DATA
200.000000,0.00
210.000000,17.00
225.000000,14.80
The choices for the 1 of N fields in the metadata are as follows:
Frequency Unit: Hz, kHz, MHz, GHz
Antenna Unit: dBuV/m, dBuA/m, dBG, dBpT, None
Frequency Interpolation: Logarithmic, Linear
Spectrum Analyzer Mode User's &Programmer's Reference 2616