Meas
138
Absolute Sets the limit test to show Fail if one offset is
larger than the corresponding limit for Abs Start
and/or Abs Stop. This is the default selection for
each offset for BTS measurements.
Relative Sets the limit test to show Fail if one offsets
spectrum emission mask measurement results is
larger than the corresponding limit for Rel Start
and/or Rel Stop.
Abs AND Rel Set the limit test to show Fail if one offset’s
spectrum emission mask measurement results is
larger than both the corresponding limit for Abs
Start and Abs Stop and the limit for Rel Start and
Rel Stop.
Abs OR Rel Set the limit test to show Fail if one offset’s the
absolute spectrum emission mask measurement
results is larger than the corresponding limit
for Abs Start and Abs Stop or the limit for Rel
Start and Rel Stop.
Meas BW Specifies the bandwidth to use when measuring the offset.
When using the front panel, only the currently selected
offset is affected. All the offsets are affected when remote
control. This is the multiplier applied to the resolution
bandwidth to determine the bandwidth to integrate when
calculating the power. Meas BW is now restricted to be 1
and not changeable.