EasyManua.ls Logo

Keysight N9320B - Improving Frequency Resolution and Accuracy

Keysight N9320B
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Functions and Measurements
Improving Frequency Resolution and Accuracy
46
Improving Frequency Resolution and Accuracy
This section provides information on using the frequency
counter function to improve frequency resolution and
accuracy.
1 Press > Preset (With Preset Type of Factory)
2 Connect a cable from the front panel CAL OUT to RF IN;
Press > Alignment > Align > CAL OUT ON to toggle on and
enable the 50 MHz amplitude reference signal.
3 Press hardkey.
The analyzer will detect the signal peak and locate it to the
center of the display screen (Refer to Auto Tune” on page 94).
4 Turn the frequency counter on:
Press
> Function > Freq Counter > Freq Counter (On).
5 Move the marker, with the front- panel knob, half- way down
the skirt of the signal response.
Figure 3-14 Using Frequency Counter
Preset/
System
Preset/
System
Auto
Tune
Marker
Marker
50.032500 MHz
– 49.30 dBm
NOTE
The frequency and amplitude of the marker appears in the active function
area (this is not the counted result). The counted result appears in the
upper-right corner of the display to the right-side of Cntr1.
Counted result