Making a Reflection Calibration Measurement
Making a Reflection Calibration Measurement
The following procedure makes a reflection measurement
using a coupler or directional bridge to measure the return
loss of a filter. This example uses a 370 MHz low pass filter
as the DUT.
The calibration standard for reflection measurements is
usually a short circuit connected at the reference plane (the
point at which the device under test (DUT) is connected.)
See Figure 3-22, a short circuit has a reflection coefficient of
1 (0 dB return loss). It reflects all incident power and
provides a convenient 0 dB reference.
1 Connect the DUT to the directional bridge or coupler as
shown in Figure 3-22. Terminate the unconnected port of the
DUT.
Figure 3-22 Reflection Measurement Short Calibration Test Setup