23
Getting Started
Helpful Tips
Helpful Tips
The following contains information to help in using and
maintaining the instrument for optimum operation, including
alignment, external reference, firmware update and option
activation.
Perform a Self Alignment
The N9320B provides a self alignment function to manually
align the amplitude based on the time base. The analyzer
should warm up for approximately 30 minutes before self
alignment.
When the self aligment function is triggered, the current
measurement is interrupted and a gauge displays on the
LCD. The gauge simply indicates alignment is in process.
When the aligment is finished, the interrupted measurement
restarts.
Please refer to the operation procedures as below:
1 Connect a BNC cable with a N- BNC adapter between the
CAL OUT and RF IN front panel connectors.
2 Press > Alignment > Alignment > Align (Ext Cable) to initiate
the alignment.
Perform a Time Base Calibration
The N9320B provides a manual calibration function to
calibrate the time base. The analyzer should warm up for
approximately 30 minutes before calibration.
When the calibration function is triggered, the current
measurement is interrupted and a gauge displays on the
LCD. The gauge simply indicates calibration action rather
than calibration course.