Key Reference
MODE
142
Tracking
Peak
Activates a routine that automatically adjusts fine
frequency tracking to obtain the peak response of the
tracking generator on the spectrum analyzer display.
Tracking Peak is performed in the current resolution
bandwidth.
Key access: > Tracking Generator > More
Man Track
adj
Allows you to manually adjust the frequency of the
tracking generator oscillator using the step keys, knob,
or numeric keypad. The tracking adjust is tuned to
maximize the amplitude of the trace.
Key access:
> Tracking Generator > More
NOTE
When in the Tracking Generator mode, resolution bandwidths < 30 kHz
are not available.
MODE
MODE