14 Keysight S5040A Open RAN Studio Player and Capture Appliance User’s Guide
1 Setting up the S5040A Appliance
Environmental Characteristics
Table 4 List of environmental characteristics for the S5040A Appliance
Characteristic Rating / Range / Description
Environment Indoor use only
Ambient Temperature Operating: +5 °C to +40 °C
Non-operating: –40 °C to +70 °C
CAUTION: If storing the S5040A Appliance above +40 °C, make sure
the feet are firmly secured beneath the instrument.
Humidity Operating: maximum relative humidity 80 percent for temperatures up
to 31 °C, decreasing linearly to 50 percent relative humidity at 40 °C,
non-condensing.
Non-operating: up to 95 percent relative humidity, non-condensing,
at 40 °C; decreasing linearly to 50 percent relative humidity at 70 °C
Altitude Operating: up to 3,000 meters (9,842 feet)
Non-operating: up to 15,300 meters (50,000 feet)
Weight 17.968 lbs (8.15 kgs)
Dimensions Width = 434 mm (17.09 in)
Depth = 495 mm (19.49 in)
Height = 106 mm (4.17 in)
Safety UL 61010-1:2010/CAN/CSA C22.2 No. 61010-1-12 Amendment 1
IEC 61010-1:2017 Amendment 1
IEC 61010-2-030:2017 Amendment 1
UL 61010-2-030:2018 Amendment 1
CAN/CSA-22.2 No. 61010-2-030-17 Amendment 1
Installation Category II