2 Operation and Features
56 Keysight U2722A/U2723A User’s Guide
Memory list commands storing and restoring
Only 11 selected type of commands can be stored in the memory list. Table 2-1
summarizes the type of commands that can be saved into the memory list.
Commands in the memory list can be stored in and restored from the nonvolatile
memory. When the U2723A USB Modular Source Measure Unit is powered up,
existing commands from the memory list is restored from the nonvolatile memory
automatically.
For more information, refer to the Keysight U2722A/U2723A USB Modular Source
Measure Units Programmer’s Reference.
NOTE
After changing the measurement range command, a clamp command should be
added subsequently into the memory list.
Table 2-1 Type of commands that can be saved into the memory list
Remote SCPI command AMM command
equivalent
Description
[SOURce:]MEMory:CURRent:LIMit CurrentLimit Set the current limit
[SOURce:]MEMory:CURRent:MEASure MeasureCurrent Make a current measurement
[SOURce:]MEMory:CURRent:SOURce SourceCurrent Set the drive current
[SOURce:]MEMory:CURRent:RANGe CurrentRange Set the current range
[SOURce:]MEMory:VOLTage:LIMit VoltageLimit Set the voltage limit
[SOURce:]MEMory:VOLTage:MEASure MeasureVoltage Make a voltage measurement
[SOURce:]MEMory:VOLTage:SOURce SourceVoltage Set the drive voltage
[SOURce:]MEMory:VOLTage:RANGe VoltageRange Set the voltage range
[SOURce:]MEMory:SOURce:DELay:AUTO AutoDelay Enable (ON) or disable (OFF) the auto delay mode
[SOURce:]MEMory:SOURce:DELay
GlobalDelay Set the global delay value
LocalDelay Set the local delay value
[SOURce:]MEMory:OUTPut Output Enable (ON) or disable (OFF) the instrument output