4 Appendices
68 Keysight U2722A/U2723A User’s Guide
Appendix B: Time-out Settings
The following procedure provides the instructions on how to set the time-out
value in the Keysight IO Libraries.
1 To launch your Keysight Connection Expert, go to Start > All Programs >
Keysight IO Libraries Suite > Keysight Connection Expert.
2 On the Instrument I/O on this PC panel, select your device in the USB
expanded list and right-click.
3 Select Send Commands To This Instrument and the Keysight Interactive IO
window will appear.
4 Go to Interactive IO > Options and the Options dialog box will appear.
5 Configure the time-out value based on the table provided below.
Table 4-2 Time-out settings
Points Tint (ms) Time-out range (ms)
1 to 10
1 to 100 Std IO 5000
101 to 32767 Assign 110% of t
11 of 50
1 to 50 Std IO 5000
51 to 32767 Assign 120% of t
51 to 100
1 to 10 Std IO 5000
11 to 32767 Assign 120% of t
101 to 500
1 to 5 Std IO 5000
6 to 500 Assign 130% of t
501 to 32767 Assign 110% of t
501 to 1000
1 to 500 Assign 310% of t
501 to 32767 Assign 120% of t
1001 to 4096
1 to 500 Assign 310% of t
501 to 32767 Assign 110% of t
NOTE
Expected time, t = Points
×
Tint
–The Points is set using the SENSe:SWEep:POINts command and the Tint is set
using the SENSe:SWEep:TINTerval command.