5 Verification and Performance Tests
Input Connections
146 U3606A User’s and Service Guide
Input Connections
Test connections to the instrument are best accomplished by using the
dual banana plug with copper wire short between the two terminals for
low- thermal offset measurement.
Shielded, twisted- pair PTFE interconnecting cables of minimum length are
recommended between the calibrator and the U3606A. Cable shields
should be earth ground referenced.
This configuration is recommended for optimal noises and settling time
performance during calibration.
Zero offset verification test setup
Figure 5-1 Test setup for zero offset (short) verification
Connect the
shorting plug
to the V and LO
terminals