6 Calibration Procedures
Adjustments procedures
200 U3606A User’s and Service Guide
6 Enter the actual applied input (see “Entering adjustment values” on
page 190).
7 Press to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
8 Repeat step 3 through step 7 for each gain adjustment point shown in
Table 6- 3.
9 Verify the DC voltage gain adjustments using the DC voltage gain
verification test” on page 157.
Table 6 - 3 DC voltage gain adjustment
Adjustment item Input voltage
Short
Dual banana plug with copper wire short
between the two input terminals
100.000 mV 100 mV
1.00000 V +1 V
–1.00000 V –1 V
10.0000 V 10 V
100.000 V 100 V
1000.00 V 1000 V
OUT
SBY
CAUTION
Set the calibrator output to 0 V before disconnecting from the
instrument input terminals.