Calibration Procedures 6
Adjustments procedures
U3606A User’s and Service Guide 201
AC voltage gain adjustment procedure
Review the “Test Considerations” on page 145 and “Gain adjustment 
considerations” on page 197 sections before beginning this procedure.
1 Press   twice to enter the AC voltage gain calibration. The AC 
annunciator will be illuminated.
2 The primary display will show the uncalibrated value and the upper 
secondary display will show the reference value of the adjustment item.
3 Configure each adjustment item shown in Table 6- 4. 
4 Use   or   to select the adjustment item.
5 Apply the input signal shown in the “Input” column of Table 6- 4.
6 Enter the actual applied input (see “Entering adjustment values” on 
page 190).
7 Press   to start the adjustment. The display “CALib” in the lower 
secondary display starts flashing to indicate that the calibration is in 
progress.
• Successful completion of the adjustment is indicated by a short beep 
and the primary display briefly showing “PASS”.
• An adjustment failure is indicated by a long beep, the primary 
display showing “FAiL” and a calibration error number appearing in 
the upper secondary display. Correct the problem and repeat this 
procedure.
8 Repeat  step 3 through step 7 for each gain adjustment point shown in 
Table 6- 4.
9 Verify the AC voltage gain adjustments using the “AC voltage 
verification test” on page 158.
Always complete tests in the same order as shown in Ta b l e 6 - 4 .