2 Digital Multimeter Operation
Making Measurements
50 U3606A User’s and Service Guide
Testing continuity
1 Press again (until the continuity annunciator is shown on the
display) to make continuity measurements.
2 Connect the red and black test leads to input terminals (red) and
LO (black) respectively as shown in Figure 2- 8.
3 Probe the test points. The instrument beeps when the continuity
measurement is less than or equal to the continuity threshold.
Table 2 - 10 Continuity function summary
Item Description
Measurement method
0.83 mA ± 0.2% constant current source, open-circuit voltage limited
to < 5 V
Audible tone: Continuous beep when reading is less than threshold resistance of
10 Ω at 1.0 kΩ range
Input protection 1000 V
rms
on all ranges, < 0.3 A short circuit
Continuity annunciator