8 Characteristics and Specifications
Digital Multimeter Specifications
244 U3606A User’s and Service Guide
Digital Multimeter Specifications
Specification assumptions
Specifications stated are for 60- minutes warm- up and 5½ digit
resolution.
One- year calibration cycle, with calibration temperature of 18 °C to
20 °C.
Operating temperature at 18 °C to 28 °C (64.4 °F to 82.4 °F).
Accuracy is expressed as: ±(% of reading + % of range)
Temperature coefficient: Add [0.1 × (the applicable accuracy) / °C] for 0
°C to 18 °C and 28 °C to 55 °C.
Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for
30 °C to 55 °C.
DC specifications
Table 8 - 1 DC accuracy specifications ± (% of reading + % of range)
Function Range
[1]
Test current or
burden voltage
24 Hours
[2]
23 ºC ± 1 ºC
90 Days
23 ºC ± 5 ºC
1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
DC voltage
19.9999 mV - 0.012 + 0.04 0.015 + 0.04 0.025 + 0.04 0.0015 + 0.0005
100.000 mV - 0.012 + 0.008 0.015 + 0.008 0.025 + 0.008 0.0015 + 0.0005
1.00000 V - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005 0.0010 + 0.0005
10.0000 V - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005 0.0020 + 0.0005
100.000 V - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005 0.0015 + 0.0005
1000.00 V - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005 0.0015 + 0.0005