EasyManua.ls Logo

Keysight U3606A - Table 8-9 Square-Wave Output Specifications

Keysight U3606A
293 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Characteristics and Specifications
DC Power Supply Specifications
258 U3606A User’s and Service Guide
Square-wave output specifications
Remote sensing capability: Output can maintain specifications for
Range S1 (30 V/1 A): Up to a 0.75-volt drop per load lead.
Range S2 (8 V/3 A): Up to a 0.5-volt drop per load lead.
Voltage programming speed:
(excludes command
processing time)
Full load
Up: 300 ms
Down: 400 ms
No load
Up: 300 ms
Down: 400 ms
Table 8 - 9 Square-wave output specifications
Parameter Specifications
Amplitude accuracy
± (offset)
Range S1 0.2 V
Range S2 0.2 V
Amplitude resolution
Range S1 1 mV
Range S2 1 mV
Frequency accuracy
± (% of frequency setting + offset)
Range
(27 steps
[1]
)
0.005% + 1 count
Frequency resolution 0.01 Hz
Duty cycle accuracy
± (% of duty cycle setting)
Range
(256 steps :
0.39% to 99.60%)
0.4%
[2][3]
Duty cycle resolution 0.39%
[3]
Pulse width accuracy
[3][4]
± (offset)
Range
(256 steps:
1/frequency)
Duty cycle/frequency
Pulse width resolution Range/256
[1] Available frequencies: 0.5, 2, 5, 6, 10, 15, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600,
2400, 4800 (Hz)

Table of Contents