Calibration Procedures 6
Adjustments procedures
U3606A User’s and Service Guide 211
CC output adjustment procedure — front output terminals
Review the “Test Considerations” on page 145 and “Input Connections” on
page 146 sections before beginning this procedure.
1 Turn off the instrument and connect the front panel (red) and
(black) output terminals to the I (red) and LO (black) input
terminals.
2 Turn on the instrument. Press Current to enter the constant current
output calibration. Ensure that the S2 (8 V/3 A) range is selected (the
S2 annunciator is illuminated). Press Current again if it is not.
3 The primary display will show the uncalibrated value and the upper
secondary display will show the reference level of the adjustment item.
4 Configure each adjustment item shown in Table 6- 11.
5 Use or to select the adjustment item.
6 Press to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
NOTE
Always complete tests in the same order as shown in Ta b l e 6 - 1 1 .