EasyManua.ls Logo

Keysight U3606A - Page 275

Keysight U3606A
293 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Characteristics and Specifications 8
Digital Multimeter Specifications
U3606A User’s and Service Guide 253
Continuity
Measurement method 0.83 mA ± 0.2% constant current source
Open circuit voltage: < +5.0 V
dc
Audible tone: Continuous beep when reading is less than threshold resistance of
10 Ω at 1.0 kΩ range
Input protection: 1000 V
rms
on all ranges, < 0.3 A short circuit
Diode
Measurement method 0.83 mA ± 0.2% constant current source
Open circuit voltage: < +5.0 V
dc
Audible tone: Continuous beep when level is below +50 mV DC
Single tone for normal forward-biased diode or semiconductor
junction where 0.3 V ¤ reading ¤ 0.8 V
Input protection: 1000 V
rms
on all ranges, < 0.3 A short circuit
Capacitance
Measurement method Computed from constant current source charge time, typical 0.2 V to
1.4 V
ac
signal level
Maximum voltage at full
scale
For 1 nF to 10 μF range: < 1.5 V
For 100 μF to 10000 μF: 0.33 V
Input protection: 1000 V
rms
on all ranges, < 0.3 A short circuit
Response time: Approximately 1 s for 100 μF and ranges below 100 μF
Charge and discharge
voltage:
5 V
pp
(approximately from +3 V to –2 V)
Frequency
Measurement method: Reciprocal counting technique
Signal level: 0.2 V to 1.4 V
Input protection:
Voltage path: 1000 V
rms
on all ranges, < 0.3 A short circuit
Current path: Protected with a 3.15 A/500 V, FF fuse
Maximum display counts (excluding frequency)
5½ digit: 120,000
4½ digit: 12,000

Table of Contents