Circulation cooling is set to continuous operation and is switched
off automatically when the door is opened.
The LABEX
®
105 PRO-ACTIVE, LABEX
®
465
PRO-ACTIVE and FROSTER LABEX
®
96 PRO-
ACTIVE models do not have circulation cooling.
The condenser transfers the generated thermal energy to the
ambient air.
Depending on the model, the condenser is fitted
either on the back wall or in the machine room.
1.5.2 SuperFrost function
The unit is equipped with a SuperFrost function.
With the SuperFrost function, it is possible to cool the unit down to
the lowest possible temperature in the shortest time
Ä
Chapter 7.4
“Switching on the SuperFrost function” on page 44.
1.5.3 Defrosting
The FROSTER MED 95 PRO-ACTIVE and
FROSTER LABEX
®
96 PRO-ACTIVE models do
not have automatic defrosting. They therefore have
to be defrosted manually if large amounts of ice
form.
The unit defrosts automatically every 12 hours.
Automatic defrosting is time and temperature-controlled.
During automatic defrosting, the system ensures that the unit main-
tains the target temperature value.
Melt water is conducted into the interior melt water container.
Condenser
Fig. 6: Condenser (example)
Automatic defrosting
Product description
Unit functions > Defrosting
27.05.2021 Refrigerator/freezer LABEX® series 15