Tab. 6: Temperature warning limits for LABEX
®
and FROSTER LABEX
®
Model Lower temperature
warning limit
Target temperature Upper temperature
warning limit
LABEX
®
105 PRO-ACTIVE
LABEX
®
465 PRO-ACTIVE
+1 °C +5 °C +10 °C
LABEX
®
288 PRO-ACTIVE
LABEX
®
340 PRO-ACTIVE
LABEX
®
468 PRO-ACTIVE
LABEX
®
520 PRO-ACTIVE
LABEX
®
720 PRO-ACTIVE
LABEX
®
288 ULTIMATE
LABEX
®
340 ULTIMATE
LABEX
®
468 ULTIMATE
LABEX
®
520 ULTIMATE
LABEX
®
720 ULTIMATE
+2 °C +5 °C +8 °C
FROSTER LABEX
®
96 PRO-ACTIVE
-35 °C -20 °C -10 °C
FROSTER LABEX
®
330 PRO-ACTIVE
FROSTER LABEX
®
530 PRO-ACTIVE
FROSTER LABEX
®
730 PRO-ACTIVE
FROSTER LABEX
®
330 ULTIMATE
FROSTER LABEX
®
530 ULTIMATE
FROSTER LABEX
®
730 ULTIMATE
-55 °C -25 °C -15 °C
6.2.2.2 Displaying and changing the temperature warning limits
1. Press button [Temperature warning max.].
ð
The upper temperature warning limit is shown on the Dis-
play.
2. Press button [Temperature warning min.].
ð
The lower temperature warning limit is shown on the Dis-
play.
Temperature warning limits
Displaying the temperature warning
limits
Upper temperature warning limit
Lower temperature warning limit
Commissioning
Programming the unit > Temperature warning limits
27.05.2021 Refrigerator/freezer LABEX® series 39