9 Maintenance
Interval Maintenance work
At least every 6 months
Check the condenser
Ä
Chapter 9.1 “Safety inspection” on page 54.
9.1 Safety inspection
Perform safety inspection every two years!
The unit should be inspected by the owner in line
with DGUV regulation 3 (formerly BGV A3).
Subject the unit to a safety inspection at regular
intervals, but at least every two years.
NOTICE!
A VDE bleeder resistance measurement must be
carried out to ensure the ESD bleeder resistance
of the entire unit. The earthing screw for this meas-
urement is located on the refrigerator door and is
indicated by an earthing symbol.
The safety inspection performed by the owner contains the fol-
lowing individual inspections and their documentation:
1 - Visual inspection
2 - Function check
3 - Temperature check
4 - Temperature alarm test
5 - Inspection of the condenser
NOTICE!
If there is any doubt that the unit is in the correct
condition, immediately stop using the unit. To pre-
vent unintended use, label the unit accordingly.
Contact the service department
Ä
“Service con-
tact” on page 4.
Contents of the safety inspection