DANGER!
Danger due to electrostatic discharge!
Electrostatic charging and discharging endangers
or invalidates the function of the explosion-pro-
tected interior.
Make sure that no electrostatic discharge can take
place in the interior.
– Before working in the interior of the unit, touch
an earthed item (e.g. door handle).
– Clean the interior only with a damp cloth.
– Make sure that the chilled goods are stable and
cannot fall over.
– Always use the LABEX
®
interior fittings.
– Store chilled goods in break-proof, conductive
and closed containers.
– Use ESD protective equipment to minimise the
risk of ESD charging.
WARNING!
Danger of infection due to inadequate hygiene,
disinfection and sterilisation!
Contact with parts that have not been cleaned,
sterilised or disinfected poses the danger of infec-
tion.
– Comply with the applicable requirements for
personal hygiene, disinfection and sterilisation.
– Clean, disinfect and sterilise the unit before
filling it for the first time.
– Clean, disinfect and sterilise the unit in line with
this instruction manual during operation.
WARNING!
Danger due to escaping refrigerant!
The refrigerant used in the unit is explosive and
pressurised, and can cause serious injury if it
comes into contact with the eyes or skin.
– During transport and set-up, do not bend or
pierce the tubing and the evaporator.
– Do not damage the surface coat (scratching it
off, for example).
– Wear safety glasses and protective gloves when
handling the refrigerant circuit.
Electrostatic discharge
Infection of the user
Escaping refrigerant
Safety
Residual risks
27.05.2021Refrigerator/freezer LABEX® series24