Display Buzzer Description Measure
Unit key
required
3
Temperature alarm (low)
n The lower temperature
warning limit is reached or
exceeded (for example after
the door has been opened for
a longer period while the
cooling machine cools down
the unit).
n Safety equipment triggers,
monitoring circuit has
switched off the cooling
machine.
n View and check the tempera-
ture warning limit
Ä
Chapter
6.2.2.2 “Displaying and
changing the temperature
warning limits” on page 39.
n If necessary, correct values of
the temperature warning
limits.
n Monitor the temperature pro-
gression.
n If it does not normalise, con-
tact the service department.
n Transfer chilled goods to new
location.
3
Tab. 12: Error messages and status displays of the PC-KIT-STICK
Display Buzzer Description Measure Unit key
–
Status display:
n Copying process is running.
n XX represents the progress of
the copying process in per-
cent.
n Do not remove USB flash
drive.
–
– Status display:
n Copying process is complete.
n Remove USB flash drive.
–
–
Status display:
n Memory of the USB flash drive
is full.
n Use USB flash drive with suffi-
cient memory capacity.
–
–
Error message:
n The USB flash drive cannot be
read.
n Use a USB flash drive for-
matted in FAT-32 format.
–
–
Error message:
n USB flash drive was removed
while copying process was
ongoing.
n Remove USB flash drive and
try again after 1 minute.
–
Status displays and error messages
Error messages
27.05.2021Refrigerator/freezer LABEX® series64