ELECTRICAL SPECIFICATIONS
OTHER MEASUREMENT APPLICATIONS
Maximum Input: 100V RMS in Voltage setting,
600V DC or 600V AC RMS in all other settings
• Diode Test: 1.5 mA max, open circuit voltage 3.0V DC
• Continuity Check: Audible signal when resistance <50Ω,
max test current 0.35mA
• Low Impedence (Low Z): Input impedence >3kΩ
Max input 600V RMS
• Auto Power off: After ~30 minutes of inactivity
• Sampling Frequency: 3 samples per second
• Bar Graph: 41 segment, 10-15 samples per second
• Overload: "OL" indicated on display, 1000V RMS in voltage
settings, 600V RMS in all other settings
• Polarity: "-" on display indicates negative polarity
• Display: 3 ¾ digit, 4000 Count LCD, TRMS technology
ELECTRICAL SPECIFICATIONS
CAPACITANCE (AUTO-RANGING)
Range Resolution Accuracy
40.00nF 10pF ±(5.0% + 35 digits)
400.0nF 0.1nF
±(3.0% + 5 digits)
4.000μF 1nF
40.00μF 10nF
400.0μF 100nF
4000μF 1μF ±(5.0% + 5 digits)
Maximum Input: 600V DC or 600V AC RMS
FREQUENCY (AUTO-RANGING)
9.999Hz 0.001Hz
±(1.0% + 5 digits)
99.99Hz 0.01Hz
999.9Hz 0.1Hz
9.999kHz 1Hz
99.99kHz 10Hz
500.0kHz 100Hz
Sensitivity:
>
8V RMS
Maximum Input: 600V DC or 600V AC RMS
DUTY CYCLE
1% to 99.9%
0.1% ±(1.2% + 2 digits)
Pulse width: 0.1 – 100ms
Frequency width: 5Hz to 10kHz
Sensitivity: >8V RMS
Maximum Input: 600V DC or 600V AC RMS
TEMPERATURE
0° to 1500°F 0.1°F / 1°F ±(2.0% + 9°F)
-18° to 816°C 0.1°C / 1°C ±(2.0% + 5°C)
WARNINGS
To ensure safe operation and service of the meter, follow these
instructions. Failure to observe these warnings can result in
severe injury or death.
• Before each use verify meter operation by measuring a known
voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Use only with CAT IV rated test leads.
• Ensure meter leads are fully seated, and keep fingers away
from the metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes
are connected.
• Use caution when working with voltages above 25V AC RMS
or 60V DC. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock,
replace batteries when a low battery indicator appears.
• Do not attempt to measure resistance or continuity on a live
circuit.
• Always adhere to local and national safety codes. Use personal
protective equipment to prevent shock and arc blast injury
where hazardous live conductors are exposed.
• Meter is IP42 dust & water resistant. Following any contact
with water, thoroughly dry meter and test lead jacks prior to
subsequent use.