EasyManua.ls Logo

Klein 5000 V2 Series - D.1 Measuring Cable Insertion Loss

Klein 5000 V2 Series
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-2 APPENDIX D Different Tow Cable Lengths and the Startup.ini File
Series 5000 V2 Sonar System Operations and Maintenance Manual P/N 11214512, Rev. 04
For example, when testing a 1000-meter cable, from SonarPro the following
command set yielded error-free operation: $MG2 and $ML7. Therefore the
Startup.ini file would be adjusted to contain the following:
set PREEMPHISIS 7
set LBOGAIN 2
D.1 Measuring Cable Insertion Loss
To get an accurate assessment of your tow cable assembly, we suggest that you use
a Hewlett Packard 4194A Impedance/Gain-Phase Analyzer or similar.
A sample plot is shown in Figure D-1.
If you do not have access to a Hewlett Packard 4194A Impedance/Gain-Phase
Analyzer, you can get a close reading by using a function generator, a 50-ohm
terminator and an oscilloscope as shown in Figure D-2.
With the function generator and the oscilloscope connected to the cable, monitor
input on the oscilloscope and input a 1 volt peak (2 volt peak-to-peak) sine wave of
the specific frequency of interest. Measure the peak output at the other end of the
cable on the scope through the 50-ohm termination.
Figure D-1: Sample Plot of Tow Cable Characteristics

Table of Contents

Related product manuals