THEORY GUIDE Power Distribution
10DEC05
TG4825-1
Page
109 of 120
INTERLOCK
The INTERLOCK protects users from exposure to electrical systems by de-energizing
selected systems when the FRONT DOOR opens.
-6.5 V DC
A18
A4
A6
SLOW SCAN CONTROLLER BOARD
not used
K2
ERASE LAMPS
K4 = RELAY
+24 V DC
K3
DC MOTOR +24 V DC
K1
LASER DRIVER DIODE BOARD -24 V DC
K1
FRONT DOOR
INTERLOCK
SWITCH
+24 V DC
-24 V DC
MSC BOARD
DC voltage for
DUPLEX CAM and
CASSETTE
DRIVE MOTOR
CONTROLLER
SLOW SCAN
BOARD
+24 V DC
LAMPS
ERASE
+24 V DC
BOARD
GALVO
-24 V DC to
LASER DRIVER
PRE-REGULATOR
BOARD
PRE-REGULATOR
H194_5032HC
LASER DIODE
DRIVER BOARD
A17
BOARD
LASER DRIVER

Table of Contents